Operating the 13C UBT Rapid Analyzer: A Step-by-Step Guide

The 13C Urea Breath Test Machine from Headway is a sophisticated tool designed for the accurate diagnosis of Helicobacter pylori infections. Understanding how to operate this machine efficiently is crucial for healthcare professionals aiming to provide timely and accurate results. Below are the essential steps involved in using the 13C UBT Rapid Analyzer.

Step-by-Step Operation of the Urea Breath Test Machine

  1. Turn on the 13C UBT Device

Begin by powering on the 13C UBT device. Connect it to a computer using the USB cable, and also connect an external printer. Ensure the printer is powered on.

  1. Open the Software

Launch the 13C UBT Device HCBT-01 software on the connected computer. Enter the username and password (default is “1”), and click login. A green status bar icon indicates a successful connection.

  1. Self-Test of the 13C Detector

After powering on, the ubt machine for H. pylori will undergo a warm-up and self-test. Wait for the interface to display “System self-test is finished, please insert the sample for testing,” indicating readiness for sample collection.

  1. Sampling Process

Record the subject’s information on the air collection bag. The patient should breathe calmly, holding their breath for over 15 seconds. Collect baseline and sample gas as follows:

– Baseline Sample: Blow into a light blue bag and seal it as a bottom gas sample.

– Test Sample: After the patient drinks 80-100 ml of water and takes a urea [¹³C] capsule, wait for 30 minutes before collecting a sample in a light pink bag.

  1. Inserting the Sample

Insert the baseline gas sample into the Baseline port and the test gas sample into the Sample port of the analyzer. The status light will turn green to confirm successful insertion.

  1. Input Information and Testing

Enter the subject’s details into the analyzer and start the test by selecting the channel number. Click “Start Measurement” to initiate the automatic testing process.

  1. Print the Report

After testing, select the record for printing, then click “Report Print” to generate the test report via the external printer.

Conclusion
The 13C Urea Breath Test Machine from Headway is a vital instrument for diagnosing H. pylori infections. By following these operational steps, healthcare providers can ensure accurate and efficient testing, leading to better patient outcomes.

About Richard

Check Also

Transform Your Coffee Routine with Empstrom’s Bean to Cup Machines

For those looking to enhance their coffee routine, Empstrom is the brand to consider, especially …

Leave a Reply

Your email address will not be published. Required fields are marked *